VALIDATION THERMOCOUPLE

High Accuracy Teflon/Kapton Insulated Thermocouples

Thermocouple for Pharma/Biotech Validation Studies

Autobond

Ultra-Premium Autobond Type T Thermocouple Probes
Designed Specifically for Autoclaves to Eliminate Leakage
The Autobond wire and probes were designed specifically for Autoclave and SIP applications, to eliminate wicking of water or steam in the thermocouple probe. The wire and probes are manufactured using our Ultra-Premium grade wire providing an
accuracy of 0.25°C @ 121°C.

The Autobond probes design eliminates the clear outer Teflon jacket which is where 90% of the moisture creep occurs. The inner Teflon coating is more of a mold which fills in the inner air pockets created by the stranded wire, thus blocking any
moisture creep.

Technical Detail
Temperature range from -196°C to 200°C
Accuracy of 0.25°C @ 121°C
Ideal for wet applications
Continuous use up to 200°C with a short peak of 250°C
Available 7-strand wire
Available also with Labels
Standard length of 6-8-10-12 m
Standard length of 20-25-30-35-40-50 feet
Benefits
No need for drip cuts
No REF CHK, UNDER, OVER, ****error
Extended probe life
Eliminates moisture
Eliminates rusting out SIMS
Reduced diameter

Premium (7 Strands)

Type “T” Thermocouple Wire - 22 AWG
High accuracy thermocouple wire is prepared with the highest purity and uniformity available to the industry. Thus, it eliminates measurement error resulting from cold working and raw material inconsistencies.
Technical Detail
Temperature Range: -220⁰C to 200⁰C
Peak Rating: 250⁰C
Accuracy: ± 0.1⁰C @ 40⁰C and ± 0.25⁰C @ 121⁰C

Ultra-Premium (3 Strands)

Type “T” Thermocouple Wire - 27 AWG
Ultra-premium thermocouple wire is available for critical process validation applications such as Freezers,Incubators, Chambers, Autoclaves, Raining autoclaves, SIP and wet applications. Ultra-Premium wire is the proven standard, delivering 0.21% accuracy (4 times better than premium thermocouple)
Technical Detail
Temperature Range: -220⁰C to 200⁰C
Peak Rating: 250⁰C
Accuracy: ± 0.1⁰C @ 40⁰C and ± 0.25⁰C @ 121⁰C

Ultra-Premium (7 Strands)

Kapton Insulated Type “T” Thermocouple
Kapton insulated thermocouple wire is rated for continuous use up to 260 ֯C and maximum temperature is 350 ֯C for critical process validation applications such as Dry Heat Tunnels, Incubators and Ovens. It is NOT for wet applications.
Technical Detail
Accuracy: 1.2⁰C @ 300⁰C
Continuous Rating @ 260⁰C for 3 months
Maximum Rating @ 350⁰C for 6 days

Teflon Tip

Thermocouple’s measurement end of the sensor is encapsulated by dual wall PTFE shrink tubing (applicable to Teflon insulated TC’s only) to extend the life of sensor and to protect the sensor from

  • Grounding error
  • Preventing moisture entering through the wire
  • Oxidation

Our Thermocouple Special Features

  • Prepared in Singapore, thus “Time Saving
  • Short lead time leads to your increased productivity
  • Competitive price

Capability

Capable of producing 36 TCs (12 Each/bundle x 3 Bundles) in 2-3 business days

Standard Length

6-8-10-12 meters per TC

Applications Used...
Freezer
Chamber
Autoclave / Raining Autoclave
SIP
Dry heat tunnel
Incubator
Oven
Freeze Dryer
Liquid Nitrogen Tank

Ordering Code

[Type] – [Length] – [Qty]
Example: TT3UP-10M-36

TYPE

For Teflon TC – 3 strands, Please indicate as TT3UP

For Teflon TC – 7 strands, Please indicate as TT7P

For Kapton TC – 7 strands, Please indicate as KT7P

LENGTH

Required length of TC in meters or feet (Example: 10M or 30 Ft.)

QTY

Specify the quantity (Example: 12 or 24 or 36)